20
PT1 / PT2 / PT3 probe calibration
This procedure is used to complete the calibration of the temperature probes PT1, PT2 and
PT3. This operation must be performed whenever a temperature probe is replaced.
Press +/- to enter a 0°C value of the temperature probe shown on the label or on the wire
sheath (e.g.: if a value of 1000.2 is shown on the PT1 sensor label, press + twice to display
1000.2). Press ENTER to confirm.
Press to go back to the Technician Menu without changing the calibration.
PT1 / PT2 / PT3 correction
This procedure is used to correct the temperature value displayed during the sterilising
process. For example, if using an external sampling tool a temperature lower than the one
displayed is measured in the chamber during the sterilisation process, the temperature
may be corrected as follows:
Press +/- to enter the correction which will change the temperature shown on the display.
Press ENTER to confirm.
Press to go back to the Technician Menu without changing the correction.
Technician Menu
Summary of Contents for STERILCLAVE 18 B
Page 2: ...2...
Page 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Page 7: ...7 Graphical reference representation Model 18 Contents...
Page 8: ...8 Graphical reference representation Model 24 Contents...
Page 9: ...9 Graphical reference representation of common components 40 Contents...
Page 10: ...10...
Page 63: ...63 Diagrams and wirings...
Page 75: ...75...