APPENDIX 1: 18/24 CYCLE SET POINTS
18/24 B - BHD CYCLE SET POINTS
18 S CYCLE SET POINTS
Pre-
vacuum
pressure
(mbar)
Steam
peak
pressure
(mbar)
Vacuum
pulse
pressure
(mbar)
Sterilisation
temperature
(°C)
Sterilis
ation
temper
ature
(min)
Drying
pressure
(mbar)
Drying
max
pressure
(mbar)
Drying
length
(min)
121 UNWRAPPED
200
2300
1240
121
15
-
-
-
134 UNWRAPPED
200
2300
1240
134
4
-
-
-
134 WRAPPED
200
1980
300
134
4
200
350
20
121
HOLLOW/POROUS
150
2100
190
121
15
200
350
20
134
HOLLOW/POROUS
150
3100
190
134
4
200
350
20
134 PRION
150
3100
190
134
18
200
350
20
B&D TEST
150
3100
190
134
3,5
200
350
5
VACUUM TEST
130
-
-
-
-
-
-
-
Pre-
vacuum
pressure
(mbar)
Steam peak
pressure
(mbar)
Vacuum
pulse
pressure
(mbar)
Sterilisation
temperature
(°C)
Sterilisat
ion
tempera
ture
(min)
Drying
pressure
(mbar)
Drying
max
pressure
(mbar)
Drying
length
(min)
121 UNWRAPPED
300
2300
1240
121
15
-
-
-
134 UNWRAPPED
300
2300
1240
134
4
-
-
-
134 WRAPPED
300
1980
300
134
4
350
500
25
121
HOLLOW/POROUS
300
2100
300
121
15
350
500
25
134
HOLLOW/POROUS
300
3100
300
134
4
350
500
25
134 PRION
300
3100
300
134
18
350
500
25
B&D TEST
300
3100
300
134
3,5
350
350
5
VACUUM TEST
300
-
-
-
-
-
-
-
Appendix 1
Summary of Contents for STERILCLAVE 18 B
Page 2: ...2...
Page 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Page 7: ...7 Graphical reference representation Model 18 Contents...
Page 8: ...8 Graphical reference representation Model 24 Contents...
Page 9: ...9 Graphical reference representation of common components 40 Contents...
Page 10: ...10...
Page 63: ...63 Diagrams and wirings...
Page 75: ...75...