17
Disconnect the cable from connector CN1 of the calibration tool and connect connector
CN2 of the calibration tool (1573 Ohm sample coil, corresponding to 150°C) to calibrate
the sensor PT1 or the connector CN3 (1832 Ohm sample coil, corresponding to 220°C) to
calibrate sensors PT2 and PT3.
Wait for at least 10 seconds and press ENTER.
Press to go back to the Technician Menu without changing the calibration.
Technician Menu
Summary of Contents for STERILCLAVE 18 B
Page 2: ...2...
Page 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Page 7: ...7 Graphical reference representation Model 18 Contents...
Page 8: ...8 Graphical reference representation Model 24 Contents...
Page 9: ...9 Graphical reference representation of common components 40 Contents...
Page 10: ...10...
Page 63: ...63 Diagrams and wirings...
Page 75: ...75...