21
Continuous cycle
A CONTINUOUS CYCLE is a test cycle which includes pre-vacuum, 2 fractioned vacuum
pulses, pressurisation, sterilisation (2 minutes) and drying (3 minutes). Minimum pressure
values higher than standard cycles are reached during the vacuum phases. The purpose
of the test cycle is to run the steriliser continuously, for instance at night, to test efficacy of
the performed maintenance operations:
Press +/- to enter the number of cycles and press ENTER to confirm.
Press to go back to the Technician Menu without making changes.
CONTINUOUS CYCLE will be available in STAND BY mode if a number higher than zero is
set; press ENTER to repeat the cycle for the set number of times.
Security code
Press ENTER to set the security code to 1.
Technician Menu
Summary of Contents for STERILCLAVE 18 B
Page 2: ...2...
Page 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Page 7: ...7 Graphical reference representation Model 18 Contents...
Page 8: ...8 Graphical reference representation Model 24 Contents...
Page 9: ...9 Graphical reference representation of common components 40 Contents...
Page 10: ...10...
Page 63: ...63 Diagrams and wirings...
Page 75: ...75...