ADDITIONAL CHECKS AND SETTINGS
Display off or not clear
Proceed as follows if no writing appears on the display and/or any if the display is not very
clear:
•
Check that power socket is powered, the
power cable
03
is correctly plugged in
and connected to the steriliser.
•
Check
fuses
32
.
Check that the
main switch
13
is pressed to position “I”.
•
Check the fuses on
electronic power board
34
.
•
If the display is not clear from where you are standing, change reading position or
angle. If the display is clearer, adjust contrast by means of trimmer
PT1
starting from
your preferred reading position (see chapter DIAGRAMS AND WIRING, Logic
electronic board).
Level probe check
Level probes
43
of both tanks are simple metal rods fitted into the container to be
monitored and insulated from it. The probe circuit is closed when liquid is present.
The presence of fouling or lime scales may create electric continuity between tank metal
and level probe rod and consequently read the presence of liquid in the tank when the
tank is empty.