36
3. Replacing the PT1 temperature probe
-
Replace
chamber temperature probe
39
.
-
Select option 3 in the Technician Menu (Board calibration PT1 page 12) and proceed
with calibration.
-
Select option 3 in the Technician Menu (PT1 Probe calibration page 15) and enter the R0
value.
Adjusting pressure setting
To adjust pressure setting select option 6 from the
Technician Menu (TP Board calibration
on page 4) and follow the instructions.
Replace the transducer if the error between the pressure shown on the display and the
pressure indicated by an external calibration tool exceeds ±50 mbar.
•
If the alarm is not resolved, possible causes are:
- sudden leakage from a valve or a disconnected pipe;
- faulty chamber temperature probe
39
;
- faulty
transducer
38
;
- faulty electronic board.
Level probe alarm
This alarm indicates inconsistent operation of the level probes in the
clean water tank
10
.
For example, if the maximum level probe detects presence of liquid, minimum level probe
and tank safety probe should also detect it.
The steriliser shows this message in STAND BY mode.
•
Empty, clean and fill
clean water tank
10
: the operation must be performed at
environmental temperature.
•
Check the level probes.
•
Replace the level probe insulation (see page 29).
•
Check water quality: Excessively pure water is corrosive (it removes salts from metals)
and cannot be detected by the probes.
In particular, DO NOT use bidistilled water or
water with a conductivity of less than 1 µS/cm.
Coil alarm
This alarm indicates that the coil element circuit is either open or interrupted.
The alarm is displayed on the steriliser in any condition coil circuit is NOT detected as
closed.
Perform the following checks in the order shown:
•
Rearm
safety thermostat
17
by removing the plug and pressing the button underneath.
The thermostat may be rearmed only if
chamber
09
has cooled down. To speed up
cooling, open
door
16
, which will however be locked.
Door
16
can be opened by
inserting
unlocking tool
37
(provided to service technicians) into the special hole on the
right under the door.
Alarms
Summary of Contents for STERILCLAVE 18 B
Page 2: ...2...
Page 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Page 7: ...7 Graphical reference representation Model 18 Contents...
Page 8: ...8 Graphical reference representation Model 24 Contents...
Page 9: ...9 Graphical reference representation of common components 40 Contents...
Page 10: ...10...
Page 63: ...63 Diagrams and wirings...
Page 75: ...75...