The status of the steriliser outputs can be displayed and modified:
•
EV1 charge to which the
water inlet pump
35
is connected in parallel
(included on all models)
•
EV2 drain (included on all models)
•
EV3 3-way (not included on 6/18 S Dynamica models)
•
EV4 steam (not included)
•
EV5 drying (not included on 6/18 S Dynamica models)
•
EV6 by-pass (not included on 6/18 S Dynamica models)
•
Water pump is the
automatic water inlet pump
46
(optional; standard on
BHD models). If the steriliser is provided with connections to the SPEEDY
WATER demineralisation system, the pump is replaced by solenoid valve EV7
(see page 51)
•
Vacuum pump (not included on 6/18 S Dynamica models)
•
Door lock (included on all models)
•
The fan(s) are
fans
42
: single on 6/18 S Dynamica and 18 S; double on other
models
•
Aux 1 is not used
•
Aux 2 is not used
•
External coil for pre-heating and drying
chamber
09
•
Internal coil for steam generation
NB.
The internal coil is:
•
The external cylindrical band arranged around
chamber
09
in models 18 S
Dynamica, 18/24 S - B - BHD
•
The series of lower and upper coils external to
chamber
09
in 6 S Dynamica
The internal coil is:
•
The coil in
chamber
09
in models 18 S Dynamica, 18/24 S - B - BHD
•
The lower coil external to
chamber
09
in 6 S Dynamica
Use the
↑↓
arrows to scroll the list and highlight the required option. Press ENTER to change
the output state (ON/OFF).
Press to go back to the Technician Menu.
Technician Menu
Summary of Contents for STERILCLAVE 18 B
Page 2: ...2...
Page 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Page 7: ...7 Graphical reference representation Model 18 Contents...
Page 8: ...8 Graphical reference representation Model 24 Contents...
Page 9: ...9 Graphical reference representation of common components 40 Contents...
Page 10: ...10...
Page 63: ...63 Diagrams and wirings...
Page 75: ...75...