48
ie-ms2-MS6-10
C)
evaluate in data logger if relay contacts live
D)
evaluate internal battery
–
third value in self test must be at least 2.6 V
6.
Verification of cabling
–
check the connection quality of cables, check visually entire cable length
for damage and the route of cables for interference, especially whether some parallel power wires
are not near.
7.
Visual inspection of transducers for possible interference or water penetration.
Make a verification protocol.
8.1.4
Recommendation for service
Service of data logger is made at manufacturer or authorized partner. There is no service allowed
without authorization from manufacturer. Unauthorized encroachment leads to loss of all
warranty. The most common damage due unauthorized manipulation with input modules is
damage of the motherboard, when modules are connected in improper way.
8.1.5
Placing out of operation after end of device life
Disconnect the power cord and return data logger to the supplier or specialized company. Notice: data
logger contains back-up Lithium battery on motherboard and on each counter input module (CTU, CTK)