Menu item Information
Item Information contains another Sub-Menu. Press ENTER key
to enter sub-menu and use
,
keys to move between items.
Leave Sub-Menu by pressing MENU key.
In submenu Information it is possible to display one fixed
message after another Data Logger Model, Data Logger Name,
Serial number, Record mode (cyclic/non-cyclic), Memory
occupation, Date and time in data logger, Language .
Menu item Communication
Submenu Communication
–
enables to display and change
Setting of communication interface, Communication speed, Data
logger address in the RS485 network, data logger IP address,
Gate IP address and Net Mask. Displayed items in menu depends
on actually adjusted communication interface and optionally on
implemented HW.
Change of configuration can be selectable protected by PIN
code, entered by the user. Way of entering PIN code is specified
in
Change of data logger communication interface
Press ENTER key to enter selection of communication
interface. Use
,
keys to select required communication
interface and press ENTER key to confirm the selection.
Adjusted communication interface must correspond with
physical connection and with SW configuration. If you
choose selection Ethernet-DHCP, IP address is set and gate
address to 0.0.0.0, network mask is set to Default (0).
Change of data logger communication speed
Press ENTER key to enter selection of communication speed.
Use
,
keys to select required communication speed and
press ENTER key to confirm the selection. This selection is
not available for Ethernet.
ATTENTION
–
standard computer COM port does not
support communication speed 230 400 Bd. If data logger
supports such speed, then it can be used in USB connection.
ENTER
MENU
Information
>>>
ENTER
MENU
Communication
>>>
ENTER
MENU
Com. port:
RS232
ENTER
MENU
Com. speed
115200 Bd