8.6 Cycle Printout - Stop Button Pressed
STATIM 6000B
SDS2R100
SN 310800A03600
UNIT # :
000
WATER QUALITY
5.1uS / 3.2ppm
CYCLE NUMBER
000040
12:27:05
2020/12/15
SOLID UNWRAPPED (N)
134 C/3.5min.
WARMING UP
100.8 C 102kPa
0:00
VACUUM DRAW
99.1 C 17kPa
1:11
PRESSURE PULSE
109.0 C 120kPa
1:30
VACUUM DRAW
106.0 C 29kPa
2:13
PRESSURE PULSE
109.0 C 120kPa
2:28
VACUUM DRAW
110.5 C 30kPa
3:10
PRESSURE PULSE
113.3 C 121kPa
3:26
CYCLE ABORTED
3:36
STOP BUTTON PRESSED
NOT STERILE
END TIME
3:36
Digital Signature #
B83AAEE17C3182E3
C2E5D68DCEDF23E4
6356E3B77A276BFD
0545AFA4F4C52434
Acceptable Tolerances:
Sterilisation time:
“Sterilisation time” (e.g. 3.5 mins) -0/+1%
Saturated Steam Pressure:
304kPa - 341kPa for Unwrapped/Wrapped cycle (205kPa - 232kPa for
Rubber and Plastics cycle)
Sterilisation Temperature:
“Specified temp” -0/+4 (134ºC-138ºC) (121ºC -125ºC for Rubber and Plastics cycle)
*data on Cycle Printout should fall within these ranges
Cycle clock:
starting at 0:00
Temp and Press
at the start of
the cycle
Software:
SDS2R100
Cycle counter:
the number
of cycles having been run
on the unit = 40
Model:
STAT
IM
6000B
Serial Number:
310800A03600
Unit Identifier:
Autoclave has
been set up as number 000
Quality of water in reservoir
Time/Date:
12:27 pm 15th December 2020
Cycle name and Parameters:
SOLID UNWRAPPED (N)
for 134°C/3.5 minutes
Warm up complete:
Start of
the first vacuum draw is 1:11
(see cycle graph - ‘A’ phase com-
plete, start of ‘B’ phase
First Pressure Pulse:
1:30 (start of ‘C’ phase)
Cycle aborted at 3:36
Reason cycle was aborted
Digital signature of the unit