30
5.4
Emergency Drawer Opening
The STAT
IM
B is equipped with a safety mechanism that keeps the chamber locked when the unit loses
power during a cycle.
To unlock the drawer without power, follow these steps:
1.
Wearing heat-resistant gloves, pull up on the
pressure relief valve ring at the back right of
the unit to exhaust steam from the chamber.
Hold the ring open until there is no more
steam coming out of the valve.
2.
Lift the front of the unit with your left hand on
the drawer latch and your right hand under the
right front corner.
3.
Reach under the front right of the unit for the
metal tab and pull it forward while your left
hand turns the drawer latch.
4.
Set the unit back down and open the drawer.
CAUTION!
HOT STEAM POSSIBLE