31
5.5
Running a Bowie-Dick Test
The Bowie-Dick test is used to ensure proper air removal is occurring in a pre-vacuum autoclave. Complete
air removal is important because pockets of cool air remaining in the chamber can compromise sterilisation.
The Bowie-Dick test runs a cycle at 134°C / 273°F for 3.5 minutes to evaluate the correct air removal.
To perform a Bowie-Dick test, you will need a Bowie-Dick device or test pack. These are NOT supplied with
your STAT
IM
B. To perform the test, follow the instructions provided by the test pack manufacturer.
Generally, the process is as follows:
1.
Open the drawer to insert a Bowie Dick test
pack. Position pack at front left on lower tray.
2.
Close and lock the drawer.
3.
From the home screen, select the
TESTS
icon.
4.
To run a Bowie Dick Test, press the
BD icon
.
5.
Press
START NOW
and allow the test to run
to completion.