32
6.
Press the
STOP
icon to open the drawer.
7.
Follow the test manufacturer’s instructions to
interpret the test results.
8.
If the unit passes the test, it is ready for use.
If the unit fails, check the test manufacturer’s
instructions and repeat the test. If the second
test fails, contact your service technician.
Presetting your Bowie-Dick test
To schedule a Bowie-Dick test before the start of the next working day, use the delayed start function.
1.
When you have finished using your steriliser for
the day, select the
Bowie-Dick
icon.
2.
Press the
DELAYED START
button.
3.
Set the time and date you would like the unit
to start the test.
4.
Press the
DELAYED START
button.
5.
Press the
BACK
icon to reset countdown. The
LCD will remain in the delayed start mode until
the cycle begins.