52
9.8
Replacing the Drawer Seal
Frequency: Every 6 months or 1,000 cycles (whichever is first)
1.
Open the drawer and unhook the tray to slide
it into the chamber and out of the way.
2.
Remove the old seal by pulling it out of
position. Clean the drawer seal channel of
any debris.
3.
Put the new seal in place and start by pressing
it into each corner. You can wet the seal with water
to ease installation.
4.
Then work the seal into the seal channel.
9.9
Cleaning an External Water Fill Tank
Frequency: Monthly or every 160 cycles (whichever is first)
1.
Drain the external fill tank.
2.
Fill the tank with a solution of distilled water and alcohol (10%)
3.
Let the solution sit for 30 minutes.
4.
Drain the tank and discard the solution.