15
2.4 Filling Your STAT
IM
B’s Water Reservoir
When filling the reservoir, only use distilled or filtered water containing less than 5 ppm total dissolved solids
(having a conductivity of less than 10 uS /cm).
NOTE:
Impurities and additives in other water sources will cause the water quality sensor to protect the unit
by preventing a cycle from starting.
To fill the water reservoir, there are three different methods:
1.
Automatic fill using a WATER FILTRATION SYSTEM.
2.
Automatic fill using an EXTERNAL WATER TANK AND AUXILIARY PUMP.
3.
MANUAL fill.
(Default)
Automatic Fill
If you are connecting your STAT
IM
B to an external filling device such as a water filtration system or external
water tank and auxiliary pump, make sure your unit is set to the AUTOMATIC filling mode. Your STAT
IM
B’s
default filling option is MANUAL.
To change this setting, follow these steps: