40
7.2
Setting Drying Time
Use this setting to lengthen or shorten drying times on selected cycles. The default drying times for each
cycle are pre-set to provide optimal drying of a standard load. Loads must always be checked for dryness.
To change drying times, select SETTINGS then CYCLES and follow these steps:
7.3
Setting the Stand-By Mode
Using this setting will reduce the warm-up time between cycles by keeping the chamber at a temperature
that is optimal for your office’s level of use.
›
STAND-BY LOW:
For low to average use. Provides a balance between keeping the chamber at 70°C
(158°F) and using a minimum of electricity.
›
STAND-BY HIGH:
For high use. Optimizes your STAT
IM
B for speed by keeping the chamber at
120°C (248°F). This is the unit’s default setting.
›
STAND-BY OFF:
For infrequent use. In this setting, the wait time will be longer (up to 12 minutes from
a cold start).
To change this setting and to modify the amount of time the unit is in Stand-by, select SETTINGS
then SYSTEM and follow these steps:
1.
Press stand-by mode.
2.
Select how warm you would like to keep the
chamber (high, low or off) as well as a start
and end time and which days of the week you
would like the schedule to be active.