Important Information About Using Your STATCLAVE
Intended Use
The STATCLAVE G4 is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use
by a health care provider to sterilize medical products by means of pressurized steam.
It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized
by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio-medical waste
or materials not compatible with steam sterilization. The processing of such loads may result in
incomplete sterilization and / or damage to the autoclave.
Please refer to the table below for program description, cycle times, temperature and dry times.
Program
Load Description
Sterilization
Temperature /
Exposure Time
Drying Time
Maximum
Load
Solid/Unwrapped
IUSS CYCLE* for
unwrapped solid
instruments (mirrors,
explorers), hinged
instruments (hemostats)
on trays.
132˚C /270˚F
for 4 minutes
Not applicable
6 kg /13.2 lbs
Solid/Wrapped
Wrapped IMS cassettes
with solid instruments,
rigid sterilization
containers with
solid instruments or
single pouched solid
instruments on a pouch
rack.
132˚C /270˚F
for 4 minutes
25 minutes (up to 6
kg /13.2 lbs of load)
6 kg /13.2 lbs
15 minutes (up to 1.4
kg / 3 lbs of load**)
Hollow/Unwrapped
IUSS CYCLE* for
unwrapped dental
handpieces on trays.
132˚C /270˚F
for 4 minutes
Not applicable
6 kg /13.2 lbs
Hollow/Wrapped
Single-pouched dental
handpieces on a pouch
rack.
132˚C /270˚F
for 4 minutes
25 minutes (up to 6
kg /13.2 lbs of load)
6 kg /13.2 lbs
15 minutes (up to 1.4
kg / 3 lbs of load)
Textiles/Porous
Wrapped
Textiles
32˚C /270˚F
for 4 minutes
20 minutes
2 kg /4.4 lbs
Rubber & Plastic
IUSS CYCLE* for
unwrapped solid or hollow
instrument.
121 ˚C /250˚F
for 20 minutes
Not applicable
2 kg /4.4 lbs
*Immediate Use Steam Sterilization cycle.
**For optimal drying of these loads, SciCan recommends using only one pouch rack loaded on the second from top rail.