background image

60

APPENDIX A

Software Icons

USB icon variations:

Door Lock icon variations:

Settings

No Internet or network connection. Press to view detail.

Network but no Internet connection. Press to view detail.

Network and Internet connected. Press to view detail.

USB key not inserted.

USB key inserted.

USB engaged and recording cycle information. Do not remove.

Door is unlocked.

Door is locked. PRESS to unlock.

PRESS to access Settings menu.

Network connection

Summary of Contents for SciCan STATCLAVE G4

Page 1: ...95 115012 US EN R1 STATCLAVE G4 Operator s Manual Copyright 2019 SciCan Ltd All rights reserved STATCLAVE G4 Chamber Autoclave OPERATOR S MANUAL...

Page 2: ...IN SECTION 7 5 Insert the load More information IN SECTION 5 4 Pull up on the door latch to open the door 2 Ensure BOTH reservoirs are filled with high quality distilled water More information IN SECT...

Page 3: ...bility with User ID PIN and Process Enforced Usage 39 8 2 Setting Drying Time 41 8 3 Setting the Stand By Mode 41 8 4 Setting Up and Using Your STATCLAVE Web Portal 41 8 5 Registering for STATCLAVE On...

Page 4: ...Chamber rack 1 Memory stick Operator s manual Power cord 1 Tray extractor with door unlocking pin 2 Pouched instrument racks 2 Long silicone tubes with elbow connectors 1 Reservoir drain tube for use...

Page 5: ...rs explorers hinged instruments hemostats on trays 132 C 270 F for 4 minutes Not applicable 6 kg 13 2 lbs Solid Wrapped Wrapped IMS cassettes with solid instruments rigid sterilization containers with...

Page 6: ...or maintain your STATCLAVE SciCan shall not be liable for incidental special or consequential damages caused by any maintenance or services performed on the STATCLAVE by a third party or for the use o...

Page 7: ...ling time and ensures the load is dry the moment you open the door Why Connect It The STATCLAVE s WiFi enabled G4 technology records and monitors information on every cycle to protect your office and...

Page 8: ...lance color touchscreen Offers real time monitoring of all the important sterilization parameters Easy to use emergency door opening Simple procedure to open the door and retrieve instruments in the e...

Page 9: ...lter bacteria retentive air filter 7 USB port 1 4 Unit Overview 8 Touchscreen 9 Chamber rack 10 Power cord input 11 Ethernet port not visible 12 Automatic fill port for clean water reservoir 13 Conden...

Page 10: ...yed start Start 2 Press START 3 Cycle in progress Current chamber temperature Current chamber pressure Stop Appears during STERILIZATION and DRYING to show time remaining Total time elapsed in the cyc...

Page 11: ...he Door The STATCLAVE will lock the chamber door when you select a cycle To unlock the door go to the home screen and follow these steps 1 PRESS lock icon 2 PRESS unlock 3 PRESS back 4 Lock icon now c...

Page 12: ...th front legs fully extended 19 5 495 mm Width 17 75 450 mm Depth 25 635 mm Weight empty 136 lbs 61 7kg Weight with full reservoirs and full load 175 lbs 79 5kg Temperature and Humidity Avoid installi...

Page 13: ...Watts for 120V 2 3 Connecting Your STATCLAVE s Water Draining System The STATCLAVE uses water from the Venturi reservoir to generate vacuum draws at the beginning and end of each cycle For the unit t...

Page 14: ...nd pull out the plug 2 Insert one elbow connector with silicone tube to the metal exhaust port A 3 Insert the other elbow connector with silicone tube to the plastic reservoir overflow drain port B 4...

Page 15: ...3 Connect the reservoir overflow tube B to a port on the waste bottle cap C 2 4 Connecting your STATCLAVE to a Network Connecting to a Wired Network IMPORTANT Tubes should not be kinked bent or otherw...

Page 16: ...from the default value and update it on a regular basis as per user s office IT policy See the web Portal Help tab for instructions on changing your password Change the network name SSID from the def...

Page 17: ...ar 3 BEEPs For more fill options see Section 4 Filling the Water Reservoirs If you do not wish to connect your STATCLAVE at this time Press and select a language Then press and to scroll to the end of...

Page 18: ...USB storage device into the USB port and make sure the bacteriological filter is secured in place 6 Check your national and local guidelines for any additional protocols and tests required before usi...

Page 19: ...or external water tank and auxiliary pump make sure your unit is set to the AUTOMATIC filling mode To change this setting from the home screen select then and then follow these steps To fill the clean...

Page 20: ...door open access the LCD home screen and select any cycle to view the pre cycle information screen 4 The clean water level check mark will turn yellow when the minimum water level is reached and green...

Page 21: ...he front 4 Hold or place the manual fill container at a level higher than the unit s fill port and loosen the cap on the container to allow water to flow into the tank 5 With the door open access the...

Page 22: ...not be sharply bent crushed or obstructed in any way 4 Open the valve on the water filtration system C to fill the accumulation tank 5 Open the accumulation tank s valve B to facilitate the flow of w...

Page 23: ...in the external tank and the auxiliary water pump should not run dry To use this method you will need the STATCLAVE automatic fill pump sold as an accessory and an external tank with a minimum diamete...

Page 24: ...nts before loading them into the autoclave Disinfectant residues and solid debris may inhibit sterilization and damage both the instruments and the STATCLAVE Lubricated instruments must be wiped thoro...

Page 25: ...ntil it locks into position at the back The front of the chamber rack should be flush with the chamber flange Using the Rack 1 Tray Configuration 2 Pouch Configuration 3 Cassette Configuration IMPORTA...

Page 26: ...h the paper side towards the tall support on the pouch rack This will optimize drying 2 Arrange pouches 2 per row to a maximum of 10 pouches to a rack 3 Load one pouch rack on the bottom rail and the...

Page 27: ...assettes to ensure air flow around cassettes and to promote proper drying For wrapped loads place a chemical indicator inside each of the wrappings Inserting drying plates 1 Remove the wire trays from...

Page 28: ...ization wrap them according to the instrument manufacturer s instructions select the appropriate wrapped cycle and allow it to run to completion DO use the trays provided with your unit to hold unwrap...

Page 29: ...hese substances can damage the tray supports trays and any metal instruments that may be present in the sterilization chamber 5 6 Using Biological and Chemical Indicators Use chemical process monitors...

Page 30: ...are full You will not be able to start a cycle if the reservoir levels are below the minimum fill lines 3 Open the door The LCD should display the UNLOCKED icon Pull up on the handle to disengage the...

Page 31: ...the icon to see the cycle name and parameters 7 Press start Allow the cycle to run to completion If the chamber is cold warming up can take up to 12 minutes If User PIN ID has been enabled you will be...

Page 32: ...a NOT STERILE message the contents are not sterile See Section 11 Troubleshooting for more information Water droplets and visible signs of moisture on sterile packaging or the tape used to secure it...

Page 33: ...the Start Delay 1 2 3 4 Enter the number of hours and minutes for the start delay Press to begin the countdown Press to stop and reset countdown If Process Enforced usage is enabled a PIN screen will...

Page 34: ...thout power follow these steps 1 Remove the emergency door unlocking pin located in the handle of the tray extractor supplied with your STATCLAVE 2 Insert the emergency door unlocking pin into the sma...

Page 35: ...d start function 4 2 3 1 At the end of the working day select the vacuum test Enter the number of hours and minutes for the start delay Press to begin the countdown Press to CANCEL Press Start Delay I...

Page 36: ...These are NOT supplied with your STATCLAVE To perform the test follow the instructions provided by the test manufacturer Generally the process is as follows If the unit passes the test it is ready fo...

Page 37: ...apped Wrapped IMS cassettes with solid instruments rigid sterilization containers with solid instruments or single pouched solid instruments on a pouch rack Maximum Load 6 Kg 13 2 lbs 132 C 270 F for...

Page 38: ...U BUTTONS SUBMENU BUTTONS WHAT TO DO WITH IT General Time Enter values Date Enter values Set Time Zone Select zone Time Update Select automatic or manual Time 12 24 Select 12 hour or 24 hour format Da...

Page 39: ...nd cycle data and unit errors to SciCan It will also receive automatic software updates to the user interface Disagree Your unit will NOT send any cycle information but it will receive automatic softw...

Page 40: ...PINs must first be assigned To set up a User ID and PIN select then and follow these steps To turn Process Enforced Usage ON or OFF or to activate DOCUMENTATION mode select then and follow these steps...

Page 41: ...whether a biological indicator and chemical indicator are included At the end of the cycle you will be prompted to report whether the indicators have passed and whether the load is dry as applicable...

Page 42: ...use Select if you are using your STATCLAVE infrequently In this setting the wait time will be longer up to 12 minutes from a cold start To change this setting and to modify the amount of time the unit...

Page 43: ...Registering for STATCLAVE Online Access From the Home screen select then and follow these steps Press the connectivity icon The connectivity screen displays information about your STATCLAVE s Interne...

Page 44: ...ing your STATCLAVE to a network 9 3 Retrieving Cycle Information Using the USB Data Back Up The USB storage device can be used to transfer cycle information stored in the unit to a computer Best pract...

Page 45: ...rinter From the home screen select then and follow these steps 9 5 Adjusting your Print Settings The STATCLAVE allows for several printer adjustments To access these settings select then and follow th...

Page 46: ...r Epson TM U220D C31C515603 CR LF 9600 248 0xF8 Under normal storage conditions a thermal document will remain legible for a minimum of 5 years Normal storage conditions of thermal documents include a...

Page 47: ...to perform any maintenance What you should do Daily Wipe the door gasket clean with a damp lint free cloth Clean external surfaces with a damp lint free cloth Weekly Clean the chamber and if applicab...

Page 48: ...t CAUTION Hot Chamber The unit s Stand by mode is preset to maintain a hot chamber Turn the unit off and allow adequate time for it to cool before performing maintenance CAUTION Hot Chamber The unit s...

Page 49: ...rners and loosen any deposits 2 Wipe the reservoir s surfaces using OPTIM wipes or a clean cotton cloth dampened with water Rinse the reservoir s surfaces with distilled water and drain it using the f...

Page 50: ...h debris to slow chamber draining and effect performance To clean or replace the filters follow these steps 1 Remove the 3 chamber filters by unscrewing them from the chamber Clean each filter under r...

Page 51: ...of any debris 2 Put the new door gasket in place and press the gasket into its seat Start at the top then bottom then sides With four sides seated push the remaining gasket completely into its seat CA...

Page 52: ...rity token to the person requesting access To obtain this code from the home screen select then and follow these steps DON T press DISABLE until the session is complete or until otherwise instructed T...

Page 53: ...tem using the reservoir draining procedure described in Section 10 6 Draining the Reservoirs for Cleaning In addition the back panel condenser must also be drained To drain the Condenser Access the ba...

Page 54: ...was in operation Wait a few minutes and try to run a cycle again before proceeding to the next solution NOTE STATCLAVE units connected to the Internet and registered with SciCan will automatically sen...

Page 55: ...ng with a cold chamber From a cold start the unit s total cycle time can take as much as 12 additional minutes Reduce the warm up time between cycles or set the unit to warm up at a specific time in t...

Page 56: ...d by Mode Make sure the chamber is cooled to room temperature before attempting a second Vacuum Test If subsequent Vacuum tests consistently fail contact your SciCan dealer Bowie Dick test has failed...

Page 57: ...auxiliary pump to see that it is correctly connected See Section 4 4 Automatic Filling Using External Water Tank and Auxiliary Pump Check if the unit is set to automatic filling but there is no pump...

Page 58: ...c 01 115488S Drain tube with quick connect to drain reservoirs 01 115480S Filter for reservoirs 1 01 115486S Mounting feet fixed and adjustable 4 pcs 01 115489S Drain bottle kit 01 115490S Exhaust tub...

Page 59: ...re returned to SciCan prepaid This warranty shall be considered to be validated if the product is accompanied by the original purchase invoice from the authorized SciCan dealer and such invoice identi...

Page 60: ...80 mm Right Side 2 50 mm Left Side 0 0 cm Front to open door 16 410 mm Back 0 0 mm Water quality 6 4 ppm 10 S cm conductivity at 25 C 77 F Minimum distilled water volume required for cycle 0 26 US gal...

Page 61: ...ess to view detail Network but no Internet connection Press to view detail Network and Internet connected Press to view detail USB key not inserted USB key inserted USB engaged and recording cycle inf...

Page 62: ...ectivity information PRESS to move back one screen Connectivity screen PRESS to access network and Internet information USB screen PRESS to view cycle history Last 5 successful cycles PRESS for detail...

Page 63: ...lock screen PRESS to view door and lock status Door CLOSED not latched Door OPEN Door CLOSED and LOCKED Chamber pressure is out of range Door cannot be unlocked at this time Door CLOSED and UNLOCKED P...

Page 64: ...OM Item National Vulnerability Database Reference Control Rationale Microcontroller logic board Vybrid VF6xx CVE 2017 7936 ERR010872 Secure Boot Vulnerability when using the Serial Downloader CVE 2017...

Reviews: