10.10 Replacing the Bacteriological Filter
Frequency: Every 6 months or 500 cycles.
1.
Open the unit door.
2.
Unscrew the bacteriological filter.
3.
Replace it with a new filter, hand tighten.
10.11 Replacing the Door Seal
Frequency: Every 6 months or 500 cycles.
1.
Remove the old gasket by pulling it out of
position. Clean the door gasket seat of any
debris.
2.
Put the new door gasket in place and press
the gasket into its seat. Start at the top, then
bottom, then sides. With four sides seated,
push the remaining gasket completely into its
seat.
CAUTION!
A bacteriological filter must always be in place during a cycle. Running a cycle without
a bacteriological filter in place will compromise the sterility of the load.
CAUTION! Hot Chamber.
The unit’s Stand-by mode is preset to maintain a hot chamber. Turn the unit off and
allow adequate time for it to cool before performing maintenance.