Sterilizers Bravo G4
- TECHNICAL MANUAL
92
TECHNICAL SHEETS
EN
The latter, once activated, is screwed to the threaded bushing fitted on the porthole support crossbar, making the
porthole move until the door is closed and then locked.
To ensure a silent rotation and a better resistance to effort, the shaft rotates on a ball slewing ring positioned inside
the front support of the servomotor unit.
The servomotor unit activation is controlled by the door itself, by means of an actuator pin which acts on the door ajar
microswitch.
In this first closing phase, the servomotor unit, supported by the sliding plate in anti-friction material, reacts
mechanically moving forward and sliding on the steel pins. The closing phase ends when the end-of-stroke
microswitch is activated and stops the servomotor.
When the cycle starts, the gearmotor is activated; it stops when a value ranging from 3.2 to 3.6A is reached.
Upon the first vacuum pulse, with a negative pressure of -0.80 bar, the door is further closed by approx. ½ of the
shaft rotation, ensuring a perfect sealing of the closure.
The opening command allows the servomotor to move back, freeing the door, which is moved away from the
chamber thanks to the thrust of the actuator pin; the end-of-stroke microswitch and the door ajar microswitch are
disabled in this phase.
SERVICE OPERATIONS
WARNING:
Before performing any type of operation on internal components, turn off the autoclave
and disconnect the mains cable.
Replacing the gearmotor
-
With the door open, remove the Benzing ring from the rear side to unlock the gearmotor of the drive shaft.
-
Remove the gearmotor carefully, making it slide towards the rear part, without removing the front retaining ring
and the pre-load spring washer.
-
Check the correct positioning of the front retaining ring and of the pre-load washer and insert the new gearmotor
in the drive shaft.
-
Exert a moderate pressure on the gearmotor to facilitate the insertion of the Benzing ring.
Replacing shaft / worm screw
-
With the door open, remove the Benzing ring from the rear side to unlock the gearmotor of the drive shaft.
-
Remove the gearmotor carefully, making it slide towards the rear part, removing the front retaining ring and the
pre-load spring washer.
-
Remove the 4 screws securing the full shaft support plate to the servomotor support.
For an easier operation in models equipped with HX10P3 vacuum pump, extract the pump from the
autoclave.
-
Slide the support out, complete with shaft/worm screw, actuator pin, plate and sliding pins, ball slewing ring.
-
Insert the new complete support; check the correct positioning of the slewing ring and of the actuator pin and lock
it using the 4 retaining screws.
-
Check the correct positioning of the front retaining ring and of the pre-load washer and insert the new gearmotor
in the drive shaft.
-
Exert moderate pressure on the gearmotor to facilitate the insertion of the Benzing ring.
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......