Sterilizers Bravo G4
- TECHNICAL MANUAL
14
FUNCTIONING
EN
4. FUNCTIONING
NOTE - UNIT OF MEASURE: Pressure (a.s.l.) 1 bar (value in relative barg at sea level)
0 relative bar = 1 absolute bar.
4.1. PRE-HEATING
When the autoclave is turned on, the pre-heating setting screen is displayed, set by default to ON with a time of 60’.
Depending on the user’s choice, it can be:
-
PRE-HEATING not active (OFF).
-
PRE-HEATING active (ON).
The pre-heating phase, depending on the user’s choice, can be modified from the menu “SETTINGS” -
“PREFERENCES” - “PRE-HEATING”.
With PRE-HEATING not active, the following components will not be powered nor activated:
-
Solenoid valves
-
Fans
-
Water pump
-
Vacuum pump
-
Chamber tube bundle heating element
-
Generator heating element
With PRE-HEATING active, the chamber tube bundle heating element is activated and managed through a
temperature control that allows keeping:
-
Chamber temperature at 50°C (122°F) with open door.
-
Chamber temperature at 100°C (212°F) with closed door.
When the device is started with PRE-HEATING active, the actual operation starts:
-
After one successfully completed cycle (test cycle or sterilization cycle).
-
After one aborted cycle (except in case of “vacuum test” cycle).
The PRE-HEATING
is disabled
:
-
When a cycle is performed with delayed start.
-
At the end of the set pre-heating time.
NOTE: The timer restarts when the door is closed when performing pre-heating with open door.
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......