Sterilizers Bravo G4
- TECHNICAL MANUAL
42
SERVICE MENU
EN
SQ11 – CHECK EV5 FOR PROPER CLOSING
Remove steel plate (
1
) to separate the door cover from the
crossbar/porthole unit
-
Be careful with the spring/cylinder units (
2
).
-
Close the door.
-
Access the SERVICE menu.
-
Select COMPONENT TEST.
From DOOR menu:
-
Lock the door.
From SOLENOID VALVES menu:
-
Open EV1 and EV4.
-
Close the other solenoid valves, if open.
From OTHER COMPONENTS menu
-
Activate VP – vacuum pump.
Then, back to the SOLENOID VALVES menu.
With EV5 closed, (not activated), there should be no air suction from the
middle of the bacteriological filter.
With EV5 open, (activated), there should be air suction from the middle of
the bacteriological filter.
Deactivate the components after completing the test.