Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
TECHNICAL MAINTENANCE
111
8.3.8. DOOR GASKET REPLACEMENT
Before proceeding, open the door then switch off the device.
Remove the old gasket from the door simply pulling it out by hands (no tools required).
Clean the door gasket seat, to ensure it is debris-free.
Install the new door gasket by pressing it into its seat, starting
from the top/bottom.
then the sides.
Finally press the remaining gasket sections completely into its
seat. Do not lubricate the gasket nor the seat.