Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
TECHNICAL SHEETS
107
Once the main board FW & connectivity module FW uploads are complete, proceed with the recovery of:
- Device serial no.
- PT1 – chamber temperature probe value.
Both required actions can be made using the dedicated controls available in the SERVICE MENU.
DEVICE SERIAL No. MEMORISATION
Refer to Chapter 5. SERVICE MENU:
- Action SQ1 - to access the Service menu.
- Action SQ3 - to enter the SN menu.
- Action SQ30 - to enter the device serial no.
Press ENTER to store the serial no, unique for of each
individual autoclave. Use the two counters to edit the
serial no., confirm with ENTER.
The first three digits are automatically assigned when
the Model reference is entered.
PT1 – CHAMBER TEMPERATURE PROBE VALUE
Refer to Chapter 5. SERVICE MENU:
- Action SQ1 - to access the Service menu.
- Action SQ3 - to access to PT1 probe calibration.
- Action SQ20 - to enter the PT1 value.
The PT1 calibration function lets you store the typical
PT1 reading, at 130.4°C (266.72°F). Use the counter
to edit the value, and confirm by pressing ENTER.
Leave the SERVICE menu.
If required, it’s possible to proceed with the COUNTER RECOVERY which allows the device cicle counter to
recovered to the total number of cycles already performed before the main PCB replacement.
COUNTER RECOVERY
Refer to Chapter 5. SERVICE MENU:
- Action SQ1 - to access the Service menu.
- Action SQ3 - to access the Cycle Counter recovery.
- Action SQ37 - to enter the cycle number.
The function COUNTER RECOVERY lets you edit the
cycle number. Confirm cycle count by pressing
ENTER.
COUNTER RECOVERY is available only when the
counter at 00000.
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...