4.8. 3rd PRESSURE RISE
Alternating vacuum decreases down to the value of - 0.80 bar with 2 pressure rises up to 1.00 bar ensures that the
residual air in the chamber is about 0.06% of the initial total amount.
Following the phases indicated above, the 3
rd
PRESSURE RISE starts, leading to the actual STERILIZATION
PROCESS, carried out at
:
•
134°C/2.15 bar (ROW versions only)
•
132°C (270°F)/2.15 bar (US/CA versions only)
•
121°C(249.8°F)/1.10 bar (ALL versions)
depending on the type of cycle selected.
In this pressure rise phase, the activation conditions of the various components are the same as the previous ones.
3
st
PRESSURE RISE - component activation status
ST
AT
U
S
EV1
High drain
(N.O.)
EV2
Air inlet
(N.C.)
EV3
Low drain
(N.C.)
EV4
Drain
deflector
(3V)
EV5
Chamber
air inlet
(N.C.)
FAN
W
AT
ER
PU
MP
- EV6
VA
CUUM
PU
MP
C
H
AMBER
T
U
BE
BUNDL
E H
EA
TI
NG
EL
EMEN
T
G
ENE
RA
TO
R
HE
AT
ING
EL
EMEN
T
Part 1
from -
°0.80
to 0.00
On
closed
On
open
Off
closed
On (1)
Off
closed
ON
ON
OFF
ON
ON
Part 2
from 0.00
to 2.14
On
closed
Off
closed
Off
closed
Off (2)
Off
closed
ON
ON
OFF
ON
ON
134°C B Cycle - 4 mins.
(ROW versions only)
132°C/270°F Hollow Wrapped Cycle - 4 mins.
(
US/CA versions only)
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......