Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
ERROR/WARNING-ALARM-HAZARD-SYSTEM ERROR
63
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
A120
PROCESS BOARD
ERROR
CONTACT TECH.
SERVICE
Reference heating
element 1k0 error.
Check R143 heating
element on the PCB.
Replace the PCB.
1
A121
PROCESS BOARD
ERROR
CONTACT TECH.
SERVICE
Reference heating
element 1k5 error.
Short-circuit R143
reference heating
element to check it.
Replace the PCB.
1
A122
PROCESS BOARD
ERROR
CONTACT TECH.
SERVICE
Reference heating
element 1k8 error.
Check R142 heating
element on the PCB.
Replace the PCB.
1
A126
WI-FI MODULE
ERROR
CONTACT TECH.
SERVICE
Communication
error with ESP32
on process board
or on L4.
Connection error with
Wi-Fi module.
Replace the PCB.
1
A127
CAN ERROR
CONTACT TECH.
SERVICE
Communication
error between
process board and
L4 via CAN.
Communication error
between graphic
interface and process
board via CAN.
Replace one PCB at a time until
the problem is solved.
1
A128
ETHERNET
MODULE ERROR
CONTACT TECH.
SERVICE
Communication
error between
graphic interface
and board.
Communication error
between graphic
interface and Ethernet
module.
Replace one PCB at a time (in
versions with two boards) until the
problem is solved.
1
A131
SOLENOID VALVE
1 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on
EV1.
Solenoid valve 1
broken.
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary.
2
PCB problem.
Replace the PCB.
A132
SOLENOID VALVE
2 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on
EV2.
Solenoid valve 2
broken.
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary.
2
PCB problem.
Replace the PCB.
A133
SOLENOID VALVE
3 ERROR
CONTACT TECH.
SERVICE
Error from solenoid
valve driver on
EV3.
Solenoid valve 3
broken.
Check the correct operation of the
solenoid valve from the Service
menu and replace it if necessary.
2
PCB problem.
Replace the PCB.
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......