Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
GUIDELINES - SAFETY TEST - VALIDATION (IF REQUIRED)
11
3.3. APPLICATION EXAMPLE WITH INSTRUMENT SECUTEST SIII+ ON STERILIZER
Measurement of ground conductor resistance.
Measurement of ground-leakage current.
Measurement of insulation resistance between network and grounding.
Disconnect power supply cable of the device and connect it to the relevant instrument socket (e.g.:
SECUTEST SIII+). Activate the device to be tested.
NOTE:
The device under test must be on and must be pre-heated in advance by operating a cycle of at
least 30 minutes, until it reaches the operating temperature, as required by the Standard.
Select the type of test with reference to standard IEC 61010, from PC or tester screen. Follow the instructions
displayed with reference to the visual check and device data entry (e.g.: serial number).
Power up the device and select the Standard.
INITIAL SETTING
CONNECTIONS WITH THE MACHINE POINTS MUST BE KEPT FOR THE TESTS DESCRIBED
BELOW.
NOTE:
Follow the instructions requested by the control instrument.
E.G.:
SECUTEST SIII+
instrument connection
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......