Sterilizers Bravo G4
- TECHNICAL MANUAL
82
ERROR/WARNING-ALARM-HAZARD-SYSTEM ERROR
EN
6.4. CODE “S” SYSTEM ERROR
(**)
1
= OK (warning) - Tap OK on pop-up window.
2
= OK + door unlock / RESET - Keep OK pressed for 3 seconds on the pop-up window and the autoclave will start
the procedure to open the door in full safety. When the RESET key appears in the left area of the LCD, hold it
pressed for 3 seconds.
CODE AND MESSAGE
ON THE DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
S001
FLASH MEMORY NOT
ACCESSIBLE
CONTACT TECH.
SERVICE
8MB Flash Memory
outside the
microswitch on the
process board
cannot be accessed.
The Flash Memory is
damaged.
Replace the PCB.
2
The PCB is damaged.
S002
FLASH MEMORY NOT
ACCESSIBLE
CONTACT TECH.
SERVICE
512MB Flash
Memory outside the
microswitch on the
process board
cannot be accessed.
The Flash Memory is
damaged.
Replace the PCB.
2
The PCB is damaged.
S005
PROBLEM WITH USB
KEY
CHANGE KEY
USB key not
accessible.
USB key is not detected.
Reconnect USB memory.
Check that the USB memory is
correctly formatted (up to 4G >
FAT/16 Kbit/sector - beyond 4G >
FAT32/16 Kbit/sector).
Replace USB memory.
2
USB card is damaged.
Replace USB card.
The PCB is damaged.
Replace the PCB.
S006
USB KEY NOT
ACCESSIBLE
CHANGE KEY
USB key not
accessible.
USB key is not detected.
Reconnect USB memory.
Check that the USB memory is
correctly formatted (up to 4G >
FAT/16 Kbit/sector - beyond 4G >
FAT32/16 Kbit/sector).
Replace USB memory.
2
USB card is damaged.
Replace USB card.
The PCB is damaged.
Replace the PCB.
S007
USB KEY FULL
CHANGE KEY
USB key full.
USB key full.
Empty USB key.
2
S009
PRINTER
DISCONNECTED
CHECK CONNECTION
Communication with
printer failed.
Printer not connected.
Check that the wiring is correct.
2
Printer not active/not
powered.
Check that the printer is turned
on/powered.
The PCB is damaged.
Replace the PCB.
S010
PRINTER PAPER OUT
CHECK PAPER
Printer: no paper.
Incorrect roll fitting.
Incorrect roll positioning.
Check that the fitting is correct.
Check that the positioning is correct.
2
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......