Sterilizers Bravo G4
- TECHNICAL MANUAL
40
SERVICE MENU
EN
SQ7 – DOOR
From the
DOOR
menu, you can establish:
-
The door status (closed/locked).
-
The status of the closing system microswitches.
-
The values of the current absorbed by the motor during the different
door closing phases.
The door can be manually unlocked, locked and fully opened.
During the door closing and locking phases, the variations of the I motor
parameter are displayed in the digital bar at the bottom of the DOOR
screen.
SQ8 – WATER
The operation of the level sensors located inside the tanks can be
checked in the
TANK MANAGEMENT
section.
Lifting a sensor, the relevant box lights up.
C indicates water conductivity value.
PT5 indicates the clear water temperature. PT5 probe integrated on water
quality sensor
SQ9 – SAFETY VALVE TEST
The
SAFETY VALVE TEST
allows checking the safety valve sealing by
means of a cycle bringing the chamber pressure over 2,64 bar.
The test does not provide a result regarding safety valve opening.
The operator who runs the test has to check the correct operation of the
valve, as in case of any other component in the Component Test menu.
The safety valve can open suddenly and remain open (for some seconds)
causing a significant steam outflow or it opens intermittently, causing only
a light blow. In both cases, the valve operates correctly, since it prevents
an anomalous chamber pressure rise.
The access to Safety Valve test requires a specific PIN. Contact
to receive the Safety Valve test PIN.
The test ends with the opening of the safety valve or at a max pressure
value of 2,80 bar. It's not possible to run more than one test in sequence.
Completed the test, the display shows the maximum chamber pressure
value, reached when safety valve open.
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......