Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
ERROR/WARNING-ALARM-HAZARD-SYSTEM ERROR
55
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
E020
DOOR CLOSING
ERROR
CONTACT TECH.
SERVICE
Door lock system
(closing) activation
timeout exceeded.
Closed door detection
microswitch fault -
microswitch check.
Check the correct operation of
the microswitch by means of the
SERVICE/COMPONENT
TEST/DOOR menu.
Replace the microswitch - see
the technical sheet ST03 -
DOOR CLOSING UNIT.
1
(then new
attempt to
turn off)
Closing system safety
pressure switch fault -
pressure switch check.
Analyse continuity between the
connection points of white
cables to check the correct
closing of the pressure switch
contact.
Replace the pressure switch –
see technical sheet ST04 -
PRESSURE TRANSDUCER -
SAFETY PRESSURE SWITCH.
Door closing
servomotor fault -
servomotor operation
check.
Check:
-
Correct operation of the
servomotor by means of the
SERVICE/COMPONENT
TEST/DOOR menu.
Replace the gearmotor - see the
technical sheet ST03 - DOOR
CLOSING UNIT.
Problem when
screwing crossbar
threaded
bushing/servomotor
worm screw - correct
screwing check.
Check:
-
Lubrication and integrity of the
threads of the 2 components.
-
Correct positioning of door
seal.
Replace:
-
Motor worm screw - see the
technical sheet ST03 - DOOR
CLOSING UNIT.
-
Crossbar threaded bushing.
E021
DOOR OPENING
ERROR
CONTACT TECH.
SERVICE
Door lock system
(opening) activation
timeout exceeded.
Door microswitch fault
- microswitch check.
Check:
-
Correct operation of the
microswitch by means of the
SERVICE/COMPONENT
TEST/DOOR menu.
Replace the microswitch.
1
(then new
attempt to
turn off)
Door closing
servomotor fault -
servomotor operation
check.
Check:
-
Correct operation of the
servomotor by means of the
SERVICE/COMPONENT
TEST/DOOR menu.
Replace the gearmotor - see the
technical sheet ST03 - DOOR
CLOSING UNIT.
Door rotating problem
when opening - hinge
check.
The door does not rotate during
the closing phase and does not
unlock the microswitch.
Check the door hinge and
lubricate it if necessary.
Summary of Contents for SciCan BRAVO G4
Page 2: ......
Page 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Page 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Page 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Page 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Page 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Page 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Page 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Page 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Page 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Page 116: ......