Sterilizers Bravo G4
- TECHNICAL MANUAL
76
ERROR/WARNING-ALARM-HAZARD-SYSTEM ERROR
EN
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
A362
PD PRESSURE
DOWNSTROKE
TIMEOUT
CONTACT TECH.
SERVICE
Vacuum phase error
during decreases in
drying cycle.
Chamber
depressurisation not
completed within the
timeout.
Clean the chamber drain filter (see
Maintenance Appendix in the User
Manual).
(**)
1
= OK (warning) - Tap OK on pop-up window.
2
= OK + door unlock / RESET - Keep OK pressed for 3 seconds on the pop-up window and the autoclave will start
the procedure to open the door in full safety. When the RESET key appears in the left area of the LCD, hold it
pressed for 3 seconds.