3. Mounting method
This unit can be used on a desk or the like, but it is constructed to be used by mounting it on a panel (instrument panel).
3-1 Selection of mounting place
3-1-1 Ambient temperature and humidity
Temperature range : 0 to 40
℃
Humidity range
: 20 to 80%RH
A stable place within the above ranges
3-1-3 Mounting angle
Front tilting angle : 0
Rear tilting angle : 0 to 15
Lateral angle
: 0
A mounting angle other than specified above
may affect the recording operation.
3-1-2 Factory environment
Select a place being separated from electric
field and magnetic field generation sources and
also free of mechanical vibrations and shocks.
Overvoltage category :
Ⅱ
(EN standard)
Pollution degree
: 2 (EN standard)
Altitude
: Lower than 2000m
Working place
: Indoors
3-2 Panel mounting method
3-2-1 External dimensions
3-2-2 Panel cutout
Panel cutout and
Minimum mounting dimensions
3-2-3 Mounting method
Insert this unit into the panel cutout of the instrument panel. Screw in four attached mounting screws into the
screw holes at 2 upper and lower places (4 places in total) on the right and left side panels of the unit. Insert the
mounted hexagonal screw heads into the round holes of the mounting brackets, press the unit toward the
instrument panel securely from the front while sliding it as illustrated below, and fasten the screws by the attached
spanner or plus screwdriver under the above condition. Be careful since the mounting brackets differ from each
other on the right and left sides (Mount the unit by 2 persons and pay attention to the injury related to dropping of
the unit.)
3-2 Panel mounting method
3-2-1 External dimensions
3-2-2 Panel cutout
Slide