12-4
■
Measuring ranges, accuracy rating, and display resolution
Kinds of inputs
Measuring ranges
Reference range
Accuracy rating
Display function
DC voltage
-10.0
to
10.0mV
±10mV
±0.05%±1digit
1μV
-20.0
to
20.0mV
±20mV
-40.0
to
40.0mV
±40mV
10μV
-80.0
to
80.0mV
±80mV
-1.25
to
1.25V
±1.25V
1mV
-2.5
to
2.5V
±2.5V
-5.0
to
5.0V
±5V
-10.0
to
10.0V
±10V
Therm
ocoupl
e
K
-200
to
500℃
±20mV
±0.05%±0.5℃
0.1℃
-200
to
900℃
±40mV
-200
to
1370℃
±80mV
±0.05%±
1
℃
E
-200
to
250℃
±20mV
±0.05%±0.7℃
-200
to
500℃
±40mV
-200
to
900℃
±80mV
±0.05%±1℃
J
-200
to
350℃
±20mV
±0.05%±0.7℃
-200
to
700℃
±40mV
-200
to
1200℃
±80mV
±0.05%±1℃
T
-200
to
400℃
±20mV
±0.05%±0.7℃
R
0
to
1760℃
±20mV
±0.05%±1℃
S
0
to
1760℃
±20mV
B
0
to
1820℃
±20mV
N
0
to
600℃
±20mV
±0.1%±0.1℃
0
to
1000℃
±40mV
0
to
1300℃
±80mV
±0.1%±1℃
W-WRe26
0
to
2315℃
±80mV
WRe5- WRe 26
0
to
2315℃
±80mV
PtRh40-PtRh20
0
to
1888℃
±20mV
NiMo-Ni
-50
to
1310℃
±80mV
Platinel
Ⅱ
0
to
500℃
±20mV
±0.1%±0.1℃
0
to
950℃
±40mV
±0.1%±1℃
0
to
1395
±80mV
U
-200
to
350℃
±20mV
±0.05%±1℃
-200
to
600℃
±40mV
L
-200
to
350℃
±20mV
-200
to
700℃
±40mV
-200
to
900℃
±80mV
Re
sist
an
ce t
hermom
ete
r
bulb
Pt100
-50
to
50℃
50Ω
±0.05%±0.3℃
0.1℃
-100
to
130℃
100Ω
-200
to
250℃
200Ω
-200
to
550℃
300Ω
JPt100
-50
to
50℃
50Ω
-100
to
130℃
100Ω
-200
to
250℃
200Ω
-200
to
550℃
300Ω
(Caution 1)
Accuracy at a room temperature of 23
℃
±2
℃
(Caution 2)
Thermocouple input does not include
any reference junction compensation accuracy
(Caution 3)
Burnout is OFF
(Caution 4)
Accuracy rating
shows % to the measuring range