13-12
13-5. Operation
Function of recording operation between several channel or operation results of time series of specific channel to
specified channel.
Data of operation of output channel is performed analog recording/digital recording/data display like data of
general input intake, and also performed recording format/skip specified.
【
Setting of output channel of operation
】
Setting parameter of input after CH037 is selected from [Unused]/[Operation]. When using for output channel of
operation, selecting [operation].
Channel parameter for operation output channel can be copied.
Refer to (
Ⅰ
. Instruction manual: 6-16, 8-17) for detail.
【
Setting method
】
1
.
Differential operation
1)
Refer to setting of differential recording and set it.
(Ⅰ.
Instruction manual
8-7.「CALC.」 )
2)
Select [SUB] or [SUI] for kind.
3)
Refer to following figure for screen of setting list display (collective 12 point).
Unrelated items for type which is set in operation output channel is not display.
4)
Refer to following figure for display of setting parameter recording.
The cursor does not move to parameter input area of unnecessary setting depending on set operation type.
【
Operation type
】
●
Inter-channel
:
Maximum, Minimum, Average, Sum
●
Time series
:
Maximum. Minimum, Average, Totalizing (1 minute to 24 hours)
●
Operation formula
:
Operation by prepared function formula
【
Output channel of operation
】
CH.No.37
-
72
Select one from following for operation
type of differential recording.
SUB
Inter-CH differential
recording
: X-Y
SUI
Standard value differential
recording
: X-C