10-1
10. Troubleshooting
If this unit does not function normally, check it according to the following table.
If the unit is still in trouble after checking it, contact your nearest sales agent or shop from which you bought
your unit.
Troubleshooting table
Symptoms
Check contents
(1) The unit does not function at all even
when turning on the power switch
.
①
Check the power terminal connections. (See 4-3-2)
②
Make sure that the power voltage is 100
-
240V AC.
(2) Abnormal noises occur when the printer
operates.
①
Check if the ribbon cassette is mounted normally. (See 5-2)
②
After opening the door, check the printer carriage for foreign
substances, and remove them, if any.
(3) Data are displayed, but not recorded at
all.
①
Make sure that recording is turned on. (REC. ON lights in the
status display section) (See 2-3)
②
Make sure that the ribbon cassette is mounted. (See 5-2)
③
Check the mounting condition of chart and chart end. (Chart End
lamp lights in the display section) (See 1-5 and 5-1)
(4) Printer operates, but the chart does not
move.
①
Feed the chart manually to make sure that the chart is fed
smooth.
②
Make sure that the chart is fed smooth by Feed key. (See -24)
(5) None of Rec key, Data P key, and
Feed key is acceptable.
Check the system setting. (See 8-16)
Key Lock lights in the display section in key lock mode.
(6) Range is set, but data are not displayed
nor recorded as analog or digital data in a
certain channel.
Check the dot setting. (See 8-4)
(7) Setting contents are confirmed to be
different even after correct setting.
When setting is entered, a setting change mark is printed.
Check the chart.
If the setting change mark is not printed, press Enter key after
high-lighting Decision of the setting item.
(8) Normal display mode is not reset even
after correct setting.
(9) Logging recording is not executed even
after setting it.
①
Set start time for logging recording is not yet.
②
Setting is not correct. (See 8-11)
③
Format [None] is set. (See 8-11)
(10)
There are times when the tag print is
out. Missing tag printing.
This unit performs priority control of dot printing. Tag printing has
relatively lower priority compared to the analog printing, in that
account, depending on the timing of internal processing, printing
may be canceled and it is not abnormal.
(11) The unit was operating normally, but
malfunction occurred abruptly.
Initialize the setting information. After restoring to normal
condition, perform resetting and check if the unit functions normally.