8-10
1
.
Operation recording
Operating conditions of peripheral units or the like are recorded at an optional position of the chart as operation
recording. This function is provided as a relative recording function between system conditions and measuring
data.
Max. 36 records can be marked on the chart.
Contact input
Data display
Digital recording
Analog recording
Communication
output
Open
OFF
OFF
Designated position (range setting)
0
Short
ON
ON
2% (5mm) to the span (right) direction
1
【
Setting method
】
1)
This operation is done by setting channel parameters. (
Ⅰ
. Instruction Manual 8-2 [CH. PARAM.] )
2)
For input, select ON/OFF (1
:
ON / 0
:
OFF).
Settable items are input chart recording lower-limit / higher-limit, unit, tag, display skip, digital printing skip,
and memory card recording skip only. The cursor also moves to the other items, but no setting can be done.
3)
Set the OFF recording position in units of % (every % step).
Set the ZERO side of the chart to 0% and the SPAN side to 100%. For this setting, input them to the chart
recording lower-limit / higher-limit setting parameter input area by numeric keys. The settable recording
position range is 0 to 99%. (The higher-limit is MAX. 100%)
4)
Unit, tag, and skip can be set in the same way as in normal input.
【
Input connection
】
【
Note
】
When the parallel scale recording of recording format is used, the recording positions
are % to the span of each area
A
B
B
The chart recording lower-limit is set to
[50] and the higher-limit is set to [52] at
the delivery time from our factory as
recording positions. By setting the chart
recording lower-limit to an optional
value, its set value +2 is automatically
expanded as the higher-limit.
Since the recording width can be
changed optionally, the chart recording
higher-limit value can be set optionally
in the same way as in the lower-limit.
If the chart recording higher-limit value
is smaller than the lower-limit value
ON/OFF recording positions are
reversed.
CH1
CH12