74
-
9. ADJUSTMENT
9.1 User Indication Adjustment /Indication Check
[Pen type]
YES
Move the pointer to 0 % position with
◄
/
►
keys.
The pointer moves to the current input value.
YES
NO
Operation mode
Press MODE switches.
Adjustment position select lamp “ZERO” lights intermittently.
Adjustment position select lamp “SPAN” lights intermittently.
“
ZERO” position: Recording 0 %
Transition to the user indication adjustment mode
“
SPAN” position: Recording 100 %
Transition to the user indication
adjustment mode
Do you want to set the
“SPAN” position, too?
The pointer moves to the current input value.
Move the pointer to 100 % position with
◄
/
►
keys.
Do you want to adjust
“ZERO” position?
NO
*For 1 pen type, pen select
lamp “1” blinks then move to
next adjustment position select
lamp.
Points where operations
are needed
Points where the instrument
operates
Select the adjustment position with
◄
/
►
keys.
Alarm lamp blinks and the adjusted value is written.
Alarm lamp blinks and the adjusted value is written.
Recording mode
Press MODE +
◄
switches.
CAL lamp blinks.
Press ENTRY key.
CAL lamp lights and the pen select lamp “1” blinks.
Press ENTRY key.
Press ENTRY key.
Press ENTRY key.
DIP Switch No 1: ON
Note: The adjustment position for the span
side cannot be selected first. The adjustment
position for the span side can be selected
after adjustment for the zero side.
The input value of adjustment position is set. set.
The input value of adjustment position is set. set.
Operation mode/lamp indication
DIP Switch No 1: OFF
Press ENTRY key.
Press ENTRY key.
Select the adjustment pen input with
◄
/
►
keys.
Press ENTRY key.
Selected pen select lamp lights
Adjustment position select lamp “ZERO” blinks
Flow Chart of User Indication Adjustment Procedure