9
Intelect Transport® Combo Therapy System
Figure 4.1
below provides physical details of the
Intelect Transport Combo. This section also
provides waveform specifications to aid in
troubleshooting.
Refer to this section when performing
troubleshooting, replacement, and repair of the
Intelect Transport Combo System.
A. Intelect Transport Combination Therapy System Physical Specifications
Dimensions
Width.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..10.125 in (25.7 cm)
Height . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 7.250 in (18.4 cm)
Depth .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 11.5 in (29.2 cm)
Weight
Standard Weight (with base) .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .5.07 lb (2.3 kg)
Battery Pack . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..1.87 lb (0.85 kg)
Power
Input. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..100 - 240 VAC, 50/60 Hz 100 VA
Output . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. +24 V, 3.125 A
Electrical Class .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. CLASS I
Electrical Type
Ultrasound TYPE B .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .
Electrotherapy TYPE BF . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .
Battery Type . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Nickel Metal Hydride (NiMH)
(1.2 V x 20 size AA)
Operating Environment
Temperature .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..Between 59 °F and 104 °F
(15° C and 40° C)
Relative Humidity .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 30%-60%
Atmospheric Pressure .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 950-1,050 h Pa
Complies with:
UL/IEC/EN 60601-1
IEC/EN 60601-1-2
IEC 60601-2-10
IEC 60601-2-5
FIGURE 4.1
4 SPECIFICATIONS
4.1 INTELECT TRANSPORT COMBO SYSTEM
WIDTH
WIDTH
HEIGHT
HEIGHT
DEPTH
DEPTH