Intelect Transport® Combo Therapy System
A. High Voltage Pulsed Current (HVPC)-
Figure 4.2 (Factory Default Setting)
The High Voltage Pulsed Current (HVPC) has
a very brief pulse duration characterized by 2
distinct peaks delivered at high voltage. The
waveform is monophasic (current flows in
one direction only). The high voltage causes a
decreased skin resistance, making the current
comfortable and easy to tolerate.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-500 V*
Polarity . . . . . . . Positive or Negative
(Negative)
Ramp . . . . . . . . .0.5 sec, 1 sec, 2 sec, 5 sec
(2 sec)
Display . . . . . . . . . . . Peak Current or Volts
(Volts)
Sweep . . . . . . . . . . . . . . . . . . . . . . . . .Off, 80/120 pps,
1/120 pps, 1/10 pps
(Off)
Frequency . . . . . . . . . . . . . . 10-120 pps
(100 pps)
Cycle Time . . . . . . 5/5, 4/12, 10/10, 10/20, 10/30,
10/50, and Continuous
(Continuous)
Constant Mode . . . . . . . . . . . . . . . . . . CV (only)**
(CV)
Treatment Time . . . . . . . . .1-60 Minutes
(20 min)
The specifications found in this section provide
the necessary waveform specifications to aid in
troubleshooting. A waveform graphic from an
oscilloscope is also provided for clarification.
Refer to this section when performing
troubleshooting, replacement, and repair of the
Intelect Transport Combo System.
FIGURE 4.2
4.2 INTELECT TRANSPORT COMBO ELECTROTHERAPY WAVEFORM SPECIFICATIONS
4 SPECIFICATIONS
* ESTI-2 Load Box
** CV = Constant Voltage