30
Intelect Transport® Combo Therapy System
6.2 THERAPY SYSTEM FAN
A. Part Number
.......................................................... 27158
B. Tools and Equipment Required
#2 Phillips Screwdriver
C. System Fan
1. Separate Top from Bottom.
Refer to 6.1,
part
C
.
2. Using a #2 Phillips Screwdriver, remove the
two Fan Retaining Screws securing the Fan
to the system Bottom.
See Figure 6.4.
3. Remove the Fan Baffle from the Fan Housing.
See Figure 6.5.
4. Replace the Fan by reversing
steps 1
through
3.
NOTE:
Do not over tighten the screws. Over
tightening will damage the threads of the
brass
standoffs.
•
FIGURE 6.4
REMOVE
SCREWS
FIGURE 6.5
REMOVE
BAFFLE
30
6 REMOVAL & REPLACEMENT
Unplug the unit from the power source before
attempting removal or replacement procedures to
prevent electrical shock.