5 TROUBLESHOOTING
B. Ultrasound Applicator Identification Test
Results
1. Unit operates as described in
steps 2, 4,
and
7.
Unit passed test.
2. "No Cal.", displays in Ultrasound channel.
a) Applicator not calibrated or needs
recalibration.
b) Possible bad Applicator. Retest with
known good Applicator.
3. Unplugged displays after ten seconds of
Applicator being connected to System.
a) Possible bad applicator. Retest with
known good Applicator.
b) Possible bad internal connection at
Ultrasound
Board.
c) Possible bad Ultrasound Board.
d) Possible bad Control Board.
5.14 ULTRASOUND APPLICATOR OUTPUT TEST
Perform this test using all available Intelect
Transport Combo Applicators used with the
System being tested.
A. Ultrasound Applicator Output Test
Procedures
1. Set up Ohmic Instruments DT 100 or
UPM DT 10 Ultrasound Power Meter per
Operator's Instructions and fill test reservoir
with
Degassed
Water.
2. Place an Applicator into the Power Meter
retainer. Make certain the Sound Head is
completely submerged in the degassed
water and centered directly over the
Stainless Steel Cone.
See Figure 5.14.
3. Zero or Tare meter.
4. Highlight Ultrasound. Press the Enter
button.
5. Highlight Duty Cycle. Press the Enter
button. Highlight Continuous and press
the Enter button.
6. Highlight Display. Press the Enter button
until
Watts
displays.
7. Press START button.
NOTE:
The position of the Sound Head over the
stainless steel cone is critical to the test results.
The Sound Head must be level and centered.