25
Intelect Transport® Combo Therapy System
8. Increase Intensity until the appropriate
Watts is displayed per
Figure 5.15.
9. Compare Power Meter readings to
Figure
5.15
to all settings for the respective
Applicator being tested as shown in
Figure 5.15.
10. Press Frequency button until 3.3 MHz is
displayed within the Frequency icon.
Repeat test and compare readings to
Figure 5.15.
NOTE:
The Applicator LED should constantly
illuminate green during the Applicator Output
tests.
B. Ultrasound Applicator Output Test Results
1. Output ranges fall within the specified
ranges as listed in
Figure 5.15.
Unit passed test.
2. Readings fall outside specified ranges of
Figure 5.15.
a) Possible bad Degassed Water in Power
Meter.
b) Possible use of Power Meter other than
Ohmic Instruments DT 100 or UPM DT 10
Ultrasound
Power
Meter.
c). Sound Head not leveled and centered
over cone in power meter.
d) Possible bad or out of calibration
Applicator.
e) Use known good Applicator.
f ) Check Ultrasound Board internal
connections.
g) Replace Ultrasound Board.
h) Replace Control Board.
FIGURE 5.15
APPLICATOR OUTPUT SPECIFICATIONS
APPLICATOR SIZE
POWER SETTING WATTS
OUTPUT RANGE
1 cm2
1*
2*
0.8 - 1.2
1.6 - 2.4
2 cm2
1
2
4
0.8 - 1.2
1.6 - 2.4
3.2 - 4.8
5 cm2
1
2
5
10
0.8 - 1.2
1.6 - 2.4
4.0 - 6.0
8.0 - 12.0
10 cm2
1
5
10
15**
20**
0.8 - 1.2
4.0 - 6.0
8.0 - 12.0
12.0 - 18.0
16.0 - 24.0
* 3.3 MHz Only
**1.0 MHz Only
5 TROUBLESHOOTING
Use only Degassed Water in Power Meter for testing
Ultrasound Applicators. Use of other types of water
will cause false test results.
Refer to page 16
for
Degassed Water Recipes.
Do not aerate water when filling Power Meter.
5.14 ULTRASOUND APPLICATOR OUTPUT
TEST CONTINUED