32
Intelect Transport® Combo Therapy System
6. Using Insulated Needle Nose Pliers,
disconnect the Power Supply Harnesses from
the Mains Connector.
See Figure 6.9.
7. Remove Power Supply from system.
8. Remove the Power Supply Shield.
9. Remove the Copper Tape from the shield.
See
Figure
6.7.
10. Replace the Power Supply by reversing
steps
1
through
5
.
NOTE:
Apply new 1/2 Copper Tape, part number
28152 to replace the tape removed in
Step 8
of this procedure.
FIGURE 6.9
6 REMOVAL & REPLACEMENT
6.3 POWER SUPPLY (CONTINUED)
Unplug the unit from the power source before
attempting removal or replacement procedures to
prevent electrical shock.
REMOVE
HARNESSES
When replacing the shield on the Power Supply.
Verify that the foil (CONDUCTIVE SIDE) of the shield
faces out and does not come in contact with the
components of the Power Supply.
RED
WIRE
GREEN W/
YELLOW STRIPE
WIRE
BLUE
WIRE