41
Intelect Transport® Combo Therapy System
7.1 CLEANING THE SYSTEM
A. Cleaning the Therapy System
With the system disconnected from the power
source, clean the system with a clean, lint free
cloth moistened with water and mild
antibacterial soap. If a more sterile cleaning is
needed, use a cloth moistened with an
antimicrobial
cleaner.
Do not submerse the system in liquids. Should
the unit accidentally become submersed,
contact the dealer or Chattanooga Group
Service Department immediately. Do not
attempt to use a system that has been wet
inside until inspected and tested by a Service
Technician Certified by Chattanooga Group.
Do not allow liquids to enter the ventilation
holes in the optional modules. This could
permanently damage the modules.
B. Cleaning Therapy System Lens
Clean the Therapy System Lens with the
NOVUS® Plastic Polishing System. NOVUS can
be purchased by going to www.novuspolish.
com on the internet. Follow the instructions as
given by NOVUS on their product.
Do Not Use alcohol or chlorine based solvents
as this may damage the lens.
7.2 CALIBRATION REQUIREMENTS
Ultrasound Applicators:
Annual calibration is required for all Ultrasound
Applicators. Only the Applicators should be sent
to the factory or a Field Technician certified
by Chattanooga Group for this procedure.
7.3 FIELD SERVICE
A. All field service procedures as described in
this Service Manual must be performed by a
Service Technician certified by Chattanooga
Group.
B. Any attempted outside the scope of this
Service Manual is the sole responsibility and
liability of the Field Technician performing
such
procedures.
C. After the performance of any Field Service,
perform the tests as described in
5.3 through
5.16
to verify the system operates properly
and within specifications prior to placing the
unit back into operation.
7.4 FACTORY SERVICE
When the Intelect Transport Combo Therapy
System requires factory service, contact
the dealer or Chattanooga Group Service
Department.
7 GENERAL MAINTENANCE