Maintenance
5-6
Replacing the Reagent Syringe Seals
NOTE
See Chapter 2, “Hardware Setup,” for an illustration of the syringe
components.
To replace the reagent syringe seals, follow these steps:
1
Remove the syringe from the pump.
2
Remove the syringe plunger from the barrel.
3
Using a single edged razor or precision knife, carefully slice the old seal lengthwise
and remove it from the plunger. Care must be taken not to damage the plunger or
“O”-rings beneath the seal.
4
Wet the “O”-ring (if present) and plunger tip with distilled or deionized water.
5
Place the seal on a flat surface with the open end facing up. Press the plunger tip
firmly into the hole until it snaps into position.
6
Lay the plunger on a flat table top, and position it so that the seal (from the “O”-ring
up) hangs over the edge.
7
Slowly roll the plunger along the table edge pressing firmly on the portion of the seal
below the “O”-ring. See Figure 5-2.
Figure 5-2. Syringe Seal Assembly
8
Rotate the plunger three complete turns. This is necessary to make the sharp raised
edge of the plunger bite into the seal for a secure fit.
9
Wet the seal with distilled or deionized water, replace the plunger, then replace the
syringe.
NOTE
Syringe sizes 250
µ
L, 500
µ
L, 1 mL, 2.5 mL, and 5.0 mL have “O”-rings.
Summary of Contents for XP 3000
Page 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Page 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Page 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Page 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Page 134: ......