3-33
Valve Leakage Detection Commands
^<n> S
ET
T
HRESHOLD
V
ALUE FOR
F
LUID
D
ETECTION
The [^] command is used to set a fluid detection value that best suits the application and
fluids used. The syntax is:
[^<n>], where <n> = 0..255
The higher the value, the more sensitive the detector. A setting of 150 detects water
leakage. Detection limits depend upon the application and the humidity of the
environment.
NOTE
Setting the value to 0 disables error reporting.