Software Communication
3-4
OEM Communication Protocol
OEM communication is a robust protocol that includes automatic recovery from
transmission errors. Table 3- describes each setting within the OEM communication
protocol.
Table 3-3. OEM Protocol (JP1-2, Jumper IN)
Parameter
Setting
Character Format
Baud rate
9600 or 38400 (set using JP1-4)
Data bits
8
Parity
None
Stop bit
1
Command Block
(see “OEM Protocol Command Block Characters”)
1
STX (^B or 02h)
2
Pump address
3
Sequence number
3+n
Data block (length n)
4+n
ETX (^C or 03h)
5+n
Checksum
Answer Block
(see “OEM Protocol Answer Block Characters”)
1
STX (^B or 02h)
2
Master address (0 or 30h)
3
Status code
3+n
Data block (length n)
4+n
ETX (^C or 03h)
5+n
Checksum
OEM P
ROTOCOL
C
OMMAND
B
LOCK
C
HARACTERS
The command block characters in the OEM communication protocol are described
below. All characters outside the command block are ignored.
When developing a parsing algorithm, the programmer should key on the STX as the
beginning of the answer block and the checksum (character after the ETX) as the end of
the answer block.
STX (^B or 02h)
The STX character indicates the beginning of a command.
Summary of Contents for XP 3000
Page 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Page 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Page 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Page 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Page 134: ......