Hardware Setup
2-18
Installing a Syringe
To install a syringe, follow these steps:
1
Loosen the plunger lock screw approximately three full turns.
2
Lower the plunger drive by sending the command [A3000R]. If power is not applied,
the plunger drive can be manually lowered by pushing down firmly on the plunger
holder assembly.
3
To install the syringe, do the following (as shown in Figure 2-11):
a
Screw the syringe into the valve.
b
Pull the syringe plunger down to the plunger holder assembly.
c
Screw the syringe plunger into place.
NOTE
Make sure the plunger lock screw is securely tightened.
Figure 2-11. Syringe Installation