Plunger Information
B-6
C
ASE
3. M
OVE TOO
S
MALL TO
R
EACH
C
UTOFF
V
ELOCITY
Case 3 is used when Vn < c
Diagram of Move:
v
V
Speed [Hz]
c
time in s
t
Vn
Calculation:
v = 50 Hz
L = 14
V = 5800 Hz
A = 5 full steps
c = 900 Hz
Theoretical Top Velocity
Vn
AL
v
=
+
=
∗ ∗ ∗
+
=
4
4 5 14 2500 50
838
2
2
Hz
if
Vn
c
<
then
Total Move Time
t
Vn
v
L
=
− =
−
∗
=
838 50
14 2500
023
.
seconds
Ramp Up Steps
A
A
1
5
= =
full steps