Software Communication
3-11
6
Select the
File
menu, and choose
Properties
. The Properties dialog box appears.
7
Select the
Settings
tab, and enter or select these options:
Function, arrow, and Control keys act as:
–
Select “Terminal keys”
Emulation:
–
Select
“Autodetect”
–
Enter “500” in Backscroll buffer lines
Click the
ASCII Setup
button. The ASCII Setup dialog box appears.
8
Enter or select these options:
–
Select “Send line ends with line feed”
–
Select “Echo typed characters locally”
–
Enter a Line delay of “0”
–
Enter a Charater delay of “0”
–
Select “Wrap lines that exceed terminal width”
9
Click
OK
to close the ASCII Setup dialog box, then click
OK
to close the Properties
dialog box.
10
Set the pump address to 0 or the appropriate address.
11
Set jumper JP1-2 to DT protocol (JP1-2 removed). Note that no jumper is needed for
microstep-enabled firmware. The communication protocol is detected automatically.
12
Power on the pump and initialize it by typing /1ZR and pressing
Enter
.
To run the pump, see the commands listed in “Using the XP 3000 Command Set” in
this chapter.